High Voltage 50 Ω Pulse Generator for CMTI Application
The CMTI-8010A is a high voltage 50 Ω pulse generator for static and dynamic common mode transfer interference (CMTI) evaluations of digital isolators according IEC-60747-5-5 and IEC-60747-17.
Features
- High voltage 50 Ω pulse generator for static and dynamic common mode transfer interference (CMTI) evaluations of digital isolators according IEC-60747-5-5 and IEC-60747-17
- 10000 V/ns CMTI slew rate at 3 kV open load output voltage and 0.3 ns rise time
- 80 kW output power into 50 Ω load
- 8 programmable pulse rise times out of: 20 ns, 10 ns, 7 ns, 5 ns, 3.5 ns, 2 ns, 1 ns, 0.6 ns, 0.3 ns. Other values on request.
- 100 ns pulse width
- Optional external pulse width extensions from 5 ns to 500 ns (up to 2 µs on request) using the external pulse width extender TLP-8012A5
- High speed 50 Ω trigger output for oscilloscopes (synchronous to high voltage pulse output)
- Built-in pulse reflection suppression
- 200 ms pulse repetition rate
- Application programming interface (API): software macros to control pulse generator in standalone operation
- Integrated interlock safety shut-down needed for pulse widths beyond 500 ns
- Industrial isolated and EMI/ESD protected USB control interface
- Two years warranty on the total system
Download: Product Datasheet CMTI-8010A
CMTI-8010A Schematic Front View
CMTI-8010A Schematic Rear View
Accessories suitcase included with CMTI-8010A (typical set)
Item | Sub-Item | Qty. | Descriptions | Model Number |
1 | 1.0 | 1 | High voltage 50 Ω pulse generator for static and dynamic common mode transfer interference evaluations of digital isolators according IEC-60747-5-5 and IEC-60747-17. An interlock safety switch is implemented inside the CMTI-8010A in order to guarantee safe usage of the high voltage capabilities. The customer has to provide adequate measurement set-up housing to guarantee the safety of the operators. | CMTI-8010A |
1.3 | 4 | BNC(m) to SMA(fm) adaptor | BNC(m)/SMA(fm) | |
1.4 | 2 | SMA(m)/SMA(m) adaptor | SMA(m)/SMA(m) | |
1.5 | 2 | SMA(f)/SMA(f) adaptor | SMA(f)/SMA(f) | |
1.6 | 2 | 50 Ω high performance SMA cable 2m, Sucoflex 104 | SF104-2m | |
1.7 | 4 | HV-BNC cables to combine pulse widths up to 1.5 µs | HV-BNC C4A Kit | |
1.8 | 1 | High voltage attenuator 50 Ω, 34 dB, 3 GHz, 2 kV | HVA-34A | |
1.9 | 1 | Transient voltage suppressor protection for scope input | TVS-70MF-10 | |
1.10 | 1 | Power line cable according supply voltage, 1.5 m | PLC-150A | |
1.11 | 1 | SMA torque wrench | 74Z-0-0-21 | |
1.12 | 1 | Hardware manual CMTI-8010A | CMTI-HM | |
1.13 | 1 | Safety switch to allow 2 µs pulse width | CMTI-IN-LO | |
1.14 | 1 | USB 2.0 Cable A Male to B Male 9.84″ (3.00m) Shielded | 102-1030-BL-F0300 |